Designing electrodes for electrical impedance spectroscopy in a four terminal setup

Main Article Content

Tobias Barth , Viviane Silva Teixeira , Wolfgang Krautschneider 

Abstract

Electrical impedance spectroscopy has a high potential for monitoring and characterizing samples, e.g., biological cells. The technique is highly sensitive but artefacts often disturb it. This study discusses problems for miniaturization of the measurement volume needed to observe single cells in a 3D printed 4T setup. There is a capacitive artefact between 1 Hz – 1 kHz introduced by the distortion of the electric field due to fringing effects. Proposed solution for reducing distortion: use of thinner measurement electrodes.

Article Details

How to Cite

Designing electrodes for electrical impedance spectroscopy in a four terminal setup. (2019). Transactions on Additive Manufacturing Meets Medicine, 1(1). https://doi.org/10.18416/AMMM.2019.1909S02T01