Kinematic Analysis of Manual, Collaborative and Robotic Surgical Retractor Holding

Main Article Content

Mathies Zehny , Jonas Meyer , Nazila Esmaeili , Jannis Hagenah 

Abstract

Surgical retractor stability is crucial for precision and safety. This study quantifies kinematic performance of manual, collaborative, and robotic retractor holding using 6-DOF tracking at 80 Hz during ex-vivo porcine liver retraction. Robotic assistance reduced position deviation from 133 mm (manual) to 0.74 mm (99.4%), tremor intensity by 99.6%, and path length from 9.3 m to 0.17 m (98%). Effect sizes ranged from d=1.03 to d=2.64. Collaborative mode achieved intermediate stability (50% deviation reduction, 93% path reduction) while preserving adjustment capability. Results establish objective metrics for retraction quality assessment and demonstrate that passive robotic assistance achieves submillimeter precision while addressing personnel shortages in open surgery.